Kategori | Multi Parameter |
Di lihat | 1404 kali |
Harga | Rp (hubungi cs) |
The 3237, 3238, and 3239 Digital HiTesters can perform 3.3 ms high-speed sampling, and come equipped with a comparator, external input and output, and an RS-232C interface. These three highperformance DMMs can be used not only in laboratories, but in production lines that require the minimal tact time. The 3239 includes the functions of the 3238 plus the 4-terminal resistance measurement function. All three units are designed with emphasis on measurement speed and safety.
In fact, HIOKI guarantees accuracy of the 3238 and 3239 for AC voltage of 10 Hz to 300 kHz, and AC current of 10 Hz to 30 kHz.
Features:
Samples at rates of up to 300 samples/ sec. ( 3.3 ms/ sample)
Comparator function provides high-speed pass/ fail evaluation
Equipped with external input and output for sequence control
Usefull Save/ Load function helps work go faster
Low power resistance measurement function prevents sample deterioration
True RMS value measurement
Interface supports full remote operation
DC V basic accuracy: ± 0.01% rdg.± 2dgt.
All the functions of the 3238, plus 4-terminal m-Ohm measurement
Basic Specifications:
DC Voltage range: 200 mV to 1000 V, 5 ranges, Best accuracy: ± 0.010 % rdg. ± 2 dgt.
AC Voltage range: 2000 mV to 700 V, 4 ranges, Best accuracy: ± 0.1 % rdg. ± 100 dgt. , Effective value rectifier
Frequency characteristics at AC Voltage: 10 to 300 kHz
Resistance range 2-terminal measurement: 200 ohm to 100 M-ohm, 7 ranges, Best accuracy: ± 0.02 % rdg. ± 2 dgt.
Resistance range 4-terminal measurement: 200 ohm to 200 k-ohm, 5 ranges, Best accuracy: ± 0.02 % rdg. ± 2 dgt.
DC Current range: 200 mA or 2000 mA, 2 ranges, Best accuracy: ± 0.1 % rdg. ± 6 dgt.
AC Current range: 200 mA or 2000 mA, 2 ranges, Best accuracy: ± 0.3 % rdg. ± 100 dgt., ( or used with clamp-on sensor range)